CVE-2025-14095 - CVE House
Back to Database
Status published Medium CVE-2025-14095

Privilege boundary violation in Radiometer Products

Vulnerability Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Impact Analysis

Refer to official advisory for detailed impact metrics.

Remediation

Ensure systems are updated to the latest vendor-supplied patch levels.

THREAT MONITOR

Am I Vulnerable?

Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2025-14095

Credits & Attribution

The following person or organization is credited with identifying this vulnerability, as recorded in the NVD database:

  • Florian Hauser and Fabian Weber from CODE WHITE GmbH

Affected Vendor

Radiometer Medical Aps

View all reports →

Affected Software

ABL90 FLEX and ABL90 FLEX PLUS Analyzers, AQT90 FLEX Analyzers, ABL800 BASIC and ABL800 FLEX Analyzers, ABL9 Analyzers
Vulnerable Versions:
All application software versions with Windows 7, Windows XP as underlying OS, Application software versions < 3.5MR11 with Windows 10 as underlying OS, All Application software versions <= 8.13 MR2, Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS, Application software versions < 6.20 MR2 with Windows 10 as underlying OS, Application software versions < 1.5.0

Timeline

Official Publish: December 17th, 2025
Last Modified: December 17th, 2025
Added to House: July 22nd, 2026

CVSS Vectors

V3: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H