Life2000 ventilator and Service PC lack sufficient audit logging capabilities
Vulnerability Description
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.
Am I Vulnerable?
Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2024-48967
Credits & Attribution
No credits recorded in the NVD database.
More from Baxter
View All →Affected Vendor
Baxter
View all reports →Affected Software
Timeline
CVSS Vectors
Weaknesses (CWE)
MITRE ATT&CK TTPs