CVE-2024-9832 - CVE House
Back to Database
Status published Critical CVE-2024-9832

No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

Vulnerability Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Impact Analysis

Refer to official advisory for detailed impact metrics.

Remediation

Ensure systems are updated to the latest vendor-supplied patch levels.

THREAT MONITOR

Am I Vulnerable?

Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2024-9832

Credits & Attribution

No credits recorded in the NVD database.

Affected Vendor

Affected Software

Life2000 Ventilation System
Vulnerable Versions:
06.08.00.00 and prior

Timeline

Official Publish: November 14th, 2024
Last Modified: November 18th, 2024
Added to House: July 22nd, 2026

CVSS Vectors

V3: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Weaknesses (CWE)