No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Vulnerability Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.
Am I Vulnerable?
Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2024-9832
Credits & Attribution
No credits recorded in the NVD database.
More from Baxter
View All →Affected Vendor
Baxter