Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Vulnerability Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.
Am I Vulnerable?
Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2025-0626
Credits & Attribution
The following person or organization is credited with identifying this vulnerability, as recorded in the NVD database:
- An anonymous researcher reported these vulnerabilities to CISA.