CVE-2020-27290 - CVE House
Back to Database
Status published Medium CVE-2020-27290

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an information...

Vulnerability Description

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an information disclosure vulnerability in the ventilator allows attackers with physical access to the configuration interface's logs to get valid checksums for tampered configuration files.

Impact Analysis

Refer to official advisory for detailed impact metrics.

Remediation

Ensure systems are updated to the latest vendor-supplied patch levels.

THREAT MONITOR

Am I Vulnerable?

Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2020-27290

Credits & Attribution

No credits recorded in the NVD database.

Affected Vendor

Affected Software

Hamilton Medical AG, T1-Ventillator
Vulnerable Versions:
Versions 2.2.3 and prior

Timeline

Official Publish: March 15th, 2021
Last Modified: August 4th, 2024
Added to House: July 21st, 2026

CVSS Vectors

V3: CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N

Weaknesses (CWE)