CVE-2020-12039 - CVE House
Back to Database
Status published Low CVE-2020-12039

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x...

Vulnerability Description

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

Impact Analysis

Refer to official advisory for detailed impact metrics.

Remediation

Ensure systems are updated to the latest vendor-supplied patch levels.

THREAT MONITOR

Am I Vulnerable?

Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2020-12039

Credits & Attribution

No credits recorded in the NVD database.

Affected Vendor

Affected Software

Baxter Sigma Spectrum Infusion Pumps
Vulnerable Versions:
Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24

Timeline

Official Publish: June 29th, 2020
Last Modified: August 4th, 2024
Added to House: July 21st, 2026

CVSS Vectors

V3: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N

Weaknesses (CWE)