Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
Vulnerability Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.
Am I Vulnerable?
Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2019-6540
Credits & Attribution
The following person or organization is credited with identifying this vulnerability, as recorded in the NVD database:
- Peter Morgan of Clever Security; Dave Singelée and Bart Preneel of KU Leuven; Eduard Marin formerly of KU Leuven, currently with University of Birmingham; Flavio D. Garcia; Tom Chothia of the University of Birmingham; and Rik Willems of University Hospital Gasthuisberg Leuven reported these vulnerabilities to CISA.
References
More from Medtronic
View All →Affected Vendor
Medtronic
View all reports →