Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
Vulnerability Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.
Am I Vulnerable?
Launch our assessment wizard to check if your infrastructure is exposed to • CVE-2019-25723
Credits & Attribution
No credits recorded in the NVD database.
References
More from Dräger
View All →Affected Vendor
Dräger
View all reports →Affected Software
Timeline
CVSS Vectors
Weaknesses (CWE)
MITRE ATT&CK TTPs
No associated TTPs found for this vulnerability.