An issue was discovered in Roche Accu-Chek Inform II Instrument...
Vulnerability Description
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.
Impact Analysis
Refer to official advisory for detailed impact metrics.
Remediation
Ensure systems are updated to the latest vendor-supplied patch levels.